← back
CVE-2020-12040CWE-319

CVE-2020-12040

3Vexday Risk Score

No sign of exploitation. No public exploitation artifact known so far.

ssvc Trackepss 0.9%
exploitation probability
0.9%top 42% of all CVEs
observed exploitation
nono source reports it
Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack.