Dräger Atlan A350 1.00 <= 1.01 DoS via Medibus Interface
13Vexday Risk Score
No sign of exploitation. No public exploitation artifact known so far.
ssvc Trackcvss 6.3epss 0.2%
exploitation probability
0.2%top 85% of all CVEs
observed exploitation
nono source reports it
Dräger Atlan A350 versions 1.00 up to and including 1.01 contains an improper input handling vulnerability that allows attackers to cause a denial of service by sending specifically crafted non-Medibus-compliant data through the Medibus interface. Attackers can transmit malformed data to overload the internal processor, gradually disrupting device operation over several hours and causing loss of data transmission, delayed display of real-time curves, and deviation between displayed airway pressure values and screen curves.
CVSS:4.0/AV:N/AC:H/AT:N/PR:N/UI:N/VC:N/VI:N/VA:L/SC:N/SI:N/SA:L
Affected products
Dräger · Atlan A350