CVE-2025-15653
Dräger Zeus IE Anesthesia Workstation USB Interface Privilege Escalation
Vexday Risk Score
21Low
SSVC decision (CISA)
Track
No exploitation signal → monitor
CVSS 7EPSS 0.2%KEV nãoPoC —Nuclei —Metasploit —Patch referenciado
Lifecycle
02 Jun 2026Published on NVD
Recommendation: Monitor — no exploitation signal at the moment.
Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.
CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Want to know if your infrastructure is exposed to this?
Talk to TrueHacking →