← volver
CVE-2018-14781mediumCWE-294

Medtronic MiniMed MMT-500/MMT-503 Remote Controllers Authentication Bypass by Capture-replay

13Vexday Risk Score

Sin señal de explotación. Ningún artefacto público de explotación conocido hasta ahora.

ssvc Trackcvss 5.3epss 0.7%
probabilidad de explotación
0.7%top 50% de las CVE
explotación observada
noninguna fuente lo reporta
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
CVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N