CVE-2015-3952
CVE-2015-3952
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.
Produtos afetados
Hospira · Plum A+3 Infusion SystemHospira · Plum A+ Infusion SystemHospira · Symbiq Infusion SystemQuer saber se a sua infraestrutura está exposta a isto?
Falar com a TrueHacking →