← voltar
CVE-2018-8868mediumCWE-749

Medtronic MyCareLink Patient Monitor Exposed Dangerous Method or Function

13Vexday Risk Score

Sem sinal de exploração. Nenhum artefato público de exploração conhecido até agora.

ssvc Trackcvss 6.2epss 0.3%
probabilidade de exploração
0.3%top 74% das CVEs
exploração observada
nãonenhuma fonte reporta
Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.
CVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L