CVE-2026-7251
Eppendorf BioFlo 320 Use of hard-coded password
Vexday Risk Score
28Low
SSVC decision (CISA)
Track
No exploitation signal → monitor
CVSS 9.3EPSS 0.5%KEV nãoPoC —Nuclei —Metasploit —Patch —
Lifecycle
26 May 2026Published on NVD
Recommendation: Monitor — no exploitation signal at the moment.
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Affected products
Eppendorf · BioFlo 320Want to know if your infrastructure is exposed to this?
Talk to TrueHacking →