CVE-2026-7251
Eppendorf BioFlo 320 Use of hard-coded password
Vexday Risk Score
28Bajo
Decisión SSVC (CISA)
Track
Sin señal de explotación → monitorear
CVSS 9.3EPSS 0.5%KEV nãoPoC —Nuclei —Metasploit —Patch —
Ciclo de vida
26 may 2026Publicada en NVD
Recomendación: Monitorear — sin señal de explotación por ahora.
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N
Productos afectados
Eppendorf · BioFlo 320¿Quieres saber si tu infraestructura está expuesta a esto?